This adjustment refreshes the default calibration used for the CAT/NA CalReady function (preset condition for the N9912A; Cal OFF is displayed).
The effects of the Error Terms adjustment are isolated to the results of the Error Terms Verification Utulity. The Error Terms adjustment does not affect the results of any other tests or adjustments.
Before performing this adjustment, it is recommended that the FieldFox port connectors be cleaned. Connection of the calibration standards should be done carefully using a torque wrench. Refer to the appropriate calibration kit user's guide for more information.
The H01/H03 calibration kits are characterized device (CD) kits with unique, precision characterization data for each short, open, and load.
|
Test Equipment |
Recommended Model Number |
|---|---|
|
Calibration kit, Type-N |
85054B-H03 |
|
Calibration kit, 3.5 mm |
85052D-H01 |
|
|