[:SENSe]:CORRection:COLLect:METHod:ERES <p1>,<p2>

(Write-Only) Sets the Cal method to Enhanced Response.

This Cal Method requires mating (insertable) DUT connectors. Use [:SENSe]:CORRection:COLLect:CONNector to change connector and gender.

Prompt for, then measure standards:

  • OPEN, SHORt, and LOAD on the <p1> (source) port.

  • THRU connection between the ports.

Learn more about FieldFox Cal Methods using SCPI.

Relevant Modes

CAT, NA

Parameters

 

<p1>,<p2>

Port numbers to calibrate.

For N9912A with Opt 110, choose 1,2.

For all other models with Full S-parameter option, choose 1,2 (forward) or 2,1 (reverse).

Examples

CORR:COLL:METH:ERES 1,2

See Cal examples

Query Syntax

Not Applicable

Default

Not Applicable


Last Modified:

18-Oct-2012

Added new models