[:SENSe]:CORRection:COLLect:METHod:SOLT2 <p1,p2>

(Write-Only) Sets the Cal Method to 2-port SOLT calibration. Calibrate for sweeps in BOTH directions.

This Cal Method is NOT available on N9912A and all other models without Opt 122.

NOTE: [:SENSe]:CORRection:COLLect:METHod:SOLR is usually a more accurate 2-port calibration. Learn more about FieldFox Cal Methods using SCPI.

Prompt for, then measure standards:

  • OPEN, SHORt, and LOAD on BOTH ports.

  • THRU connection between the ports. This should be a known (characterized) THRU standard.

Relevant Modes

CAT and NA Mode

To Cal in VVM mode, perform a Cal in CAT or NA mode, then switch to VVM mode.

Parameters

 

<p1,p2>

Port numbers to be calibrated. Choose 1,2.

Examples

CORR:COLL:METH:SOLT2 1,2

See Cal examples

Query Syntax

Not Applicable

Default

Not Applicable