CALCulate:FSIMulator: CIRCuit: | CATalog | DEVice | PORTs | [:ARRay] | REVerse | EMBED | TYPE | FILE | MODify? | PARameter | C | C2 | DELay | G | G2 | L | L2 | LOSS | VALue | R | R2 | RIN | ROUT | Z0 | STATe | VNA | PORTs | EXTension:PORT: | DELay | DISTance | VALue | UNIT | LOSS | VALue | STATe | MEDium | WAVEguide | FCUToff | COUPle | VELocity | FACTor | COUPle | STATe | POWER | COMPensate | MODE | PORT | COMPensate | STATe | SAVE | SECTion |CIRCuit | ENABle | EXTension | ENABle | ZCONversion | ENABle | TOPology | LOAD | SAVE | ZCONversion | DIFFerential | :STATe |
:BPORt |
:LPORt | COMMonmode | :STATe |
:BPORt |
:LPORt<pnum> | SENDed:PORT | :STATe | :COMPLex | :SCALar |
See Also
Example Programs
Applicable Models: E505xA with VNA ports (Write only) Copies the draft fixture to active fixture. Exact same functionality as CALC:FSIM:DRAFT:APPL |
|
Parameters |
Not applicable |
Example |
CALC:FSIM:APPL calculate:fsimulatorapply |
Return Type |
Not applicable |
Not applicable |
CALCulate<cnum>:FSIMulator:CIRCuit:CATalog?
Applicable Models: E505xA with VNA ports (Read only) Returns a comma-separated list of circuit numbers created in the draft circuit. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
Examples |
CALC:FSIM:CIRC:CAT? calculate2:fsimulator:circuit:catalog? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit:CATalog? |
Return Type |
comma-separated string |
Not Applicable |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:DEVice:PORTs[:ARRay]? <port list - in, out>
Applicable Models: E505xA with VNA ports (Read only) Returns the Device ports for the specified circuit. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:DEV:PORT? calculate2:fsimulator:circuit:device:ports? |
Query Syntax |
CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit<circN>:DEVice:PORTs[:ARRay]? |
Return Type |
Numeric |
Not Applicable |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:DEVice:PORTs:REVerse?
Applicable Models: E505xA with VNA ports (Read only) Returns whether or not the ports on an existing S2P file are reversed. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:DEV:PORT:REV? calculate2:fsimulator:circuit:device:ports:reverse? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:DEVice:PORTs:REVerse? |
Return Type |
Boolean |
OFF |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:EMBED:TYPE?
Applicable Models: E505xA with VNA ports (Read only) Returns whether the circuit is embedded or de-embedded. Valid for all block Types The block is create using CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit[n]:ADD <block type>, <fixt port count>. Setting a value not used by the selected circuit will have no affect. Learn more. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:EMBED:TYPE? calculate2:fsimulator:circuit:embed:type? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:EMBED:TYPE? |
Return Type |
character |
EMBed |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:FILE?
Applicable Models: E505xA with VNA ports (Read only) Returns the filename. Valid blocks: FILe, D4PFile, FGLoop The block is create using CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit[n]:ADD <block type>, <fixt port count>. Setting a value not used by the selected circuit will have no affect. Learn more. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:FILE? calculate2:fsimulator:circuit:file? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:FILE? |
Return Type |
string |
Not Applicable |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:FILE:EXTRapolate?
Applicable Models: E505xA with VNA ports (Read only) Returns whether or not extrapolation is allowed (TRUE/FALSE). Valid blocks: FILe, D4PFile, FGLoop. The block is create using CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit[n]:ADD <block type>, <fixt port count>. Setting a value not used by the selected circuit will have no affect. Learn more. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:FILE:EXTR? calculate2:fsimulator:circuit:file:extrapolate? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:FILE:EXTRapolate? |
Return Type |
Boolean |
OFF |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:FILE:MODify?
Applicable Models: E505xA with VNA ports (Read-only) Returns the current state of the Modify setting (NONE, NREFlect, or NXTalk). Set the Modify state using the CALCulate:FSIMulator:DRAFt:CIRCuit:FILE:MODify command. Learn more. The block is create using CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit[n]:ADD <block type>, <fixt port count>. Setting a value not used by the selected circuit will have no affect. Learn more. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:FILE:MOD? calculate2:fsimulator:circuit:file:modify? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:FILE:MODify? |
Return Type |
Enumeration |
NONE |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:C?
Applicable Models: E505xA with VNA ports (Read only) Returns the capacitance (C) circuit element for the specified circuit. Valid blocks: RLGLoop, RCGLoop, CGGLoop, SLPC, PCSL, PLSC, SCPL, PLPC, SCPC, PCSC, SLPL, PLSL. The block is create using CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit[n]:ADD <block type>, <fixt port count>. Setting a value not used by the selected circuit will have no affect. Learn more. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:PAR:C? calculate2:fsimulator:circuit:parameter:C? |
Query Syntax |
CALCulate<cnum>:FSIMulato:CIRCuit<circN>:PARameter:C? |
Return Type |
Numeric |
0 |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:C2?
Applicable Models: E505xA with VNA ports (Read only) Returns the capacitance (C2) circuit element for the specified circuit. Valid blocks: SLPC, PCSL, PLSC, SCPL, PLPC, SCPC, PCSC, SLPL, PLSL The block is create using CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit[n]:ADD <block type>, <fixt port count>. Setting a value not used by the selected circuit will have no affect. Learn more. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:PAR:C2? calculate2:fsimulator:circuit:parameter:C2? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:C2? |
Return Type |
Numeric |
0 |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:DELay ?
Applicable Models: E505xA with VNA ports (Read only) Returns port extension delay in time. Valid blocks: ILIN. The block is create using CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit[n]:ADD <block type>, <fixt port count>. Setting a value not used by the selected circuit will have no affect. Learn more. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:PAR:DEL? calculate2:fsimulator:circuit:parameter:delay? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:DEL? |
Return Type |
Numeric |
0 |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:G?
Applicable Models: E505xA with VNA ports (Read only) Returns the conductance (G) circuit element for the specified circuit. Valid blocks: RLGLoop, RCGLoop, CGGLoop, SLPC, PCSL, PLSC, SCPL, PLPC, SCPC, PCSC, SLPL, PLSL. The block is create using CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit[n]:ADD <block type>, <fixt port count>. Setting a value not used by the selected circuit will have no affect. Learn more. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:PAR:G? calculate2:fsimulator:circuit:parameter:g? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:G? |
Return Type |
Numeric |
0 |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:G2?
Applicable Models: E505xA with VNA ports (Read only) Returns the conductance (G2) circuit element for the specified circuit. Valid blocks: SLPC, PCSL, PLSC, SCPL, PLPC, SCPC, PCSC, SLPL, PLSL The block is create using CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit[n]:ADD <block type>, <fixt port count>. Setting a value not used by the selected circuit will have no affect. Learn more. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:PAR:G2? calculate2:fsimulator:circuit:parameter:g2? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:G2? |
Return Type |
Numeric |
0 |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:L?
Applicable Models: E505xA with VNA ports (Read only) Returns the inductance (L) circuit element for the specified circuit. Valid blocks: RLGLoop, RCGLoop, CGGLoop, SLPC, PCSL, PLSC, SCPL, PLPC, SCPC, PCSC, SLPL, PLSL. The block is create using CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit[n]:ADD <block type>, <fixt port count>. Setting a value not used by the selected circuit will have no affect. Learn more. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:PAR:L? calculate2:fsimulator:circuit:parameter:l? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:L? |
Return Type |
Numeric |
0 |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:L2?
Applicable Models: E505xA with VNA ports (Read only) Returns the inductance (L2) circuit element for thecspecified circuit. Valid blocks: SLPC, PCSL, PLSC, SCPL, PLPC, SCPC, PCSC, SLPL, PLSL The block is create using CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit[n]:ADD <block type>, <fixt port count>. Setting a value not used by the selected circuit will have no affect. Learn more. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:PAR:L2? calculate2:fsimulator:circuit:parameter:l2? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:L2? |
Return Type |
Numeric |
0 |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:LOSS:VALue?
Applicable Models: E505xA with VNA ports (Read only) Returns the loss at DC. Valid blocks: ILIN. The block is create using CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit[n]:ADD <block type>, <fixt port count>. Setting a value not used by the selected circuit will have no affect. Learn more. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:PAR:LOSS:VAL? calculate2:fsimulator:circuit:parameter:loss:value? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:LOSS:VALue? |
Return Type |
Numeric |
0 |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:R?
Applicable Models: E505xA with VNA ports (Read only) Returns the resistance (R) circuit element for the specified circuit. Valid blocks: RLGLoop, RCGLoop, CGGLoop, SLPC, PCSL, PLSC, SCPL, PLPC, SCPC, PCSC, SLPL, PLSL. The block is create using CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit[n]:ADD <block type>, <fixt port count>. Setting a value not used by the selected circuit will have no affect. Learn more. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:PAR:R? calculate2:fsimulator:circuit:parameter:r? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:R? |
Return Type |
Numeric |
0 |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:R2?
Applicable Models: E505xA with VNA ports (Read only) Returns the resistance (R2) circuit element for the specified circuit. Valid blocks: SLPC, PCSL, PLSC, SCPL, PLPC, SCPC, PCSC, SLPL, PLSL The block is create using CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit[n]:ADD <block type>, <fixt port count>. Setting a value not used by the selected circuit will have no affect. Learn more. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:PAR:R2? calculate2:fsimulator:circuit:parameter:r2? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:R2? |
Return Type |
Numeric |
0 |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:RIN?
Applicable Models: E505xA with VNA ports (Read only) Returns the resistance (R) value at input of transformer (real only). Valid blocks: TRANs The block is create using CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit[n]:ADD <block type>, <fixt port count>. Setting a value not used by the selected circuit will have no affect. Learn more. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:DRAF:CIRC:PAR:RIN? calculate2:fsimulator:draft:circuit:parameter:rin? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:RIN? |
Return Type |
Numeric |
0 |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:ROUT?
Applicable Models: E505xA with VNA ports (Read only) Returns the resistance (R) value at output of transformer (real only). Valid blocks: TRANs The block is create using CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit[n]:ADD <block type>, <fixt port count>. Setting a value not used by the selected circuit will have no affect. Learn more. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:PAR:ROUT? calculate2:fsimulator:circuit:parameter:rout? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:ROUT? |
Return Type |
Numeric |
0 |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:Z0?
Applicable Models: E505xA with VNA ports (Read only) Returns the Impedance (Z0) for the ideal block (scalar value). Valid blocks: ILIN. The block is create using CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit[n]:ADD <block type>, <fixt port count>. Setting a value not used by the selected circuit will have no affect. Learn more. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:PAR:Z0? calculate2:fsimulator:circuit:parameter:z0? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:PARameter:Z0? |
Return Type |
Numeric |
0 |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:STATe?
Applicable Models: E505xA with VNA ports (Read only) Returns the circuit ON/OFF. Valid for all block types The block is create using CALCulate<cnum>:FSIMulator:DRAFt:CIRCuit[n]:ADD <block type>, <fixt port count>. Setting a value not used by the selected circuit will have no affect. Learn more. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:STAT? calculate2:fsimulator:circuit:state? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:STATe? |
Return Type |
Boolean |
ON |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:VNA:PORTs? <port list>
Applicable Models: E505xA with VNA ports (Read only) Returns the VNA ports for the specified circuit. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<circN> |
Circuit Number |
Examples |
CALC:FSIM:CIRC:VNA:PORT? calculate2:fsimulator:circuit:vna:ports? |
Query Syntax |
CALCulate<cnum>:FSIMulator:CIRCuit<circN>:VNA:PORTs? |
Return Type |
Numeric |
Not Applicable |
Applicable Models: E505xA with VNA ports (Read-Write) Returns the delay value in time at the specified port. Note: This command affects ALL measurements on the specified channel. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1. |
<pnum> |
Port Number that will receive the delay setting. If unspecified, value is set to 1. |
<value> |
The port extension in seconds; may include suffix. Choose a number between: -1E18 and 1E18e. |
Examples |
CALC:FSIM:EXT:PORT1:DEL? |
Query Syntax |
CALCulate<cnum>:FSIMulator:EXTension:PORT<pnum>:DELay? |
Return Type |
Numeric |
0 |
Applicable Models: E505xA with VNA ports (Read-only) Returns the units for specifying port extension delay in physical length (distance). |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1. |
<pnum> |
Port Number. This number is ignored and all ports have the same setting. |
<char> |
Units for delay in distance. Choose from:
|
Examples |
CALC:FSIM:EXT:PORT2:DIST:UNIT? |
Query Syntax |
CALCulate<cnum>:FSIMulator:EXTension:PORT:DISTance:UNIT? |
Return Type |
Character |
METer |
Applicable Models: E505xA with VNA ports (Read-only) Returns the port extension delay in physical length (distance). |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1. |
<pnum> |
Port Number that will receive the delay setting. If unspecified, value is set to 1. |
<value> |
Physical length of fixture of added transmission line. |
Examples |
CALC:FSIM:EXT:PORT1:DIST:VAL? |
Query Syntax |
CALCulate<cnum>:FSIMulator:EXTension:PORT<pnum>:DISTance:VALue? |
Return Type |
Numeric |
0 |
Applicable Models: E505xA with VNA ports (Read-only) Returns the frequency for the Freq and Loss pair number and for the specified port number. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<pnum> |
Port Number that will receive the freq/loss settings. If unspecified, value is set to 1. |
<n> |
Freq and Loss pair number. Choose from 1 or 2. If unspecified, value is set to 1. |
<value> |
Frequency value. Choose a frequency within the frequency span of the VNA. |
Examples |
CALC:FSIM:EXT:PORT1:LOSS2:FREQ? |
Query Syntax |
CALCulate<cnum>:FSIMulator:EXTension:PORT<pnum>:LOSS<n>:FREQuency? |
Return Type |
Numeric |
0 |
Applicable Models: E505xA with VNA ports (Read-only) Returns the Loss value for the specified port number. Learn about Loss Compensation values. Note: This command affects ALL measurements on the specified channel. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<pnum> |
Port Number that will receive the Freq/Loss settings. If unspecified, value is set to 1. |
<n> |
Loss "Use" number. Choose from 1 or 2. If unspecified, value is set to 1. |
<value> |
Loss in dB. Choose a value between -90 and 90 |
Examples |
CALC:FSIM:EXT:PORT:LOSS1:VAL? calculate:fsimulator:extension:port2:loss2:value? |
Query Syntax |
CALCulate<cnum>:FSIMulator:EXTension:PORT<pnum>:LOSS<n>:VALue? |
Return Type |
Numeric |
0 |
Applicable Models: E505xA with VNA ports (Read-only) Returns the ON/OFF state for the Freq and Loss pair number and for the specified port number. Learn about Loss Compensation values. Note: This command affects ALL measurements on the specified channel. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<pnum> |
Port Number that will receive the Freq/Loss settings. If unspecified, value is set to 1. |
<n> |
Freq and Loss pair. Choose from 1 or 2. If unspecified, value is set to 1. |
<value> |
State of Freq and Loss values for port extension. 0 or OFF Specified Freq and Loss values are OFF 1 or ON Specified Freq and Loss values are ON |
Examples |
CALC:FSIM:EXT:PORT:LOSS:STAT? |
Query Syntax |
CALCulate<cnum>:FSIMulator:EXTension:PORT<pnum>:LOSS<n>[:STATe]? |
Return Type |
Boolean |
OFF |
Applicable Models: E505xA with VNA ports (Read-only) Returns the Port Loss at DC value for the specified port number. Learn about Loss Compensation values. Note: This command affects ALL measurements on the specified channel. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<pnum> |
Port number to receive Loss value. If unspecified, value is set to 1. |
<value> |
Loss in dB. Choose a value between -90 and 90 |
Examples |
CALC:FSIM:EXT:PORT:DCL:VAL? calculate:fsimulator:extension:port2:dcloss:value? |
Query Syntax |
CALCulate<cnum>:FSIMulator:EXTension:PORT<pnum>:DCLoss:VALue? |
Return Type |
Numeric |
0 |
Applicable Models: E505xA with VNA ports (Read-only) Reurns the media type of the added fixture or transmission line. See also SENS:CORR:EXT:PORT:SYSMedia Note: This command affects ALL measurements on the specified channel. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<pnum> |
Port Number. This number is ignored and all ports have the same setting. |
<char> |
Medium type. Choose from:
|
Examples |
CALC:FSIM:EXT:PORT:MED? calculate:fsimulator:extension:port2:medium? |
Query Syntax |
CALCulate<cnum>:FSIMulator:EXTension:PORT<pnum>:MEDium? |
Return Type |
Character |
COAX |
Applicable Models: E505xA with VNA ports (Read-only) Returns the cuttoff (minimum) frequency of the added waveguide fixture or transmission line. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<pnum> |
Port Number for which media type is being set. If unspecified, value is set to 1. |
<value> |
Cutoff frequency in Hz. This value is ignored when CALC:FSIM:EXT:PORT:MED is set to COAX for the same port. |
Examples |
CALC:FSIM:EXT:PORT:WAVE:FCUT? calculate:fsimulator:extension:port2:waveguide:cutoff? |
Query Syntax |
CALCulate<cnum>:FSIMulator:EXTension:PORT<pnum>:WAVEguide:FCUToff? |
Return Type |
Numeric |
0 |
Applicable Models: E505xA with VNA ports (Read-only) Returns the state of coupling with the system Velocity Factor value. Learn more. Note: This command potentially affects ALL measurements on the VNA. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<pnum> |
Port Number. This number is ignored and all ports have the same setting. |
<bool> |
Coupling state. Choose from:
|
Examples |
CALC:FSIM:EXT:PORT:WAVE:COUP? calculate:fsimulator:extension:port2:waveguide:couple? |
Query Syntax |
CALCulate<cnum>:FSIMulator:EXTension:PORT<pnum>:WAVEguide:COUPle? |
Return Type |
Boolean |
1 or ON (Coupled) |
Applicable Models: E505xA with VNA ports (Read-only) Returns the velocity factor of the fixture or added transmission line. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<pnum> |
Port Number for which velocity factor is being set. If unspecified, value is set to 1. |
<value> |
Velocity Factor. Set SENS:CORR:EXT:PORT:SYSV to use the system velocity factor. |
Examples |
CALC:FSIM:EXT:PORT:VEL:FAC? calculate:fsimulator:extension:port2:velocity:factor? |
Query Syntax |
CALCulate<cnum>:FSIMulator:EXTension:PORT<pnum>:VELocity:FACtor? |
Return Type |
Numeric |
0 |
Applicable Models: E505xA with VNA ports (Read-only) Returns the state of coupling with the system Velocity Factor value. Learn more. Note: This command potentially affects ALL measurements on the VNA. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<pnum> |
Port Number. This number is ignored and all ports have the same setting. |
<bool> |
Coupling state. Choose from:
|
Examples |
CALC:FSIM:EXT:PORT:VEL:COUP? calculate:fsimulator:extension:port2:velocity:couple? |
Query Syntax |
CALCulate<cnum>:FSIMulator:EXTension:PORT<pnum>:VELocity:COUPle? |
Return Type |
Boolean |
1 or ON (Coupled) |
Applicable Models: E505xA with VNA ports (Read-only) Returns the state of port extension for the specified port. Note: This command potentially affects ALL measurements on the VNA. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<pnum> |
Port Number for which port extension is being set. If unspecified, value is set to 1. |
<bool> |
Port extension state. Choose from:
|
Examples |
CALC:FSIM:EXT:PORT:STAT? calculate:fsimulator:extension:port2:state? |
Query Syntax |
CALCulate<cnum>:FSIMulator:EXTension:PORT<pnum>[:STATe]? |
Return Type |
Boolean |
0 |
Applicable Models: E505xA with VNA ports (Read-Wrrite) Adjust the source power compensation for gain/loss through “ALL” fixture components or through “DEEMbed” components only. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<char> |
Mode selection. Choose from:
|
Examples |
CALC:FSIM:POW:COMP:MODE ALL calculate:fsimulator:power:compensate:mode deembed |
Query Syntax |
CALCulate<cnum>:FSIMulator:POWer:COMPensate:MODE? |
Return Type |
Character |
ALL |
Applicable Models: E505xA with VNA ports (Read-Wrrite) Sets and returns on power compensation for the active fixture |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
<pnum> |
Port Number for which port extension is being set. If unspecified, value is set to 1. |
<bool> |
power compensation state. Choose from:
|
Examples |
CALC:FSIM:POW:PORT:COMP ON calculate:fsimulator:power:port:compensate:state 0 |
Query Syntax |
CALCulate<cnum>:FSIMulator:POWer:PORT[p]:COMPensate[:STATe ]? |
Return Type |
Boolean |
0 |
Applicable Models: E505xA with VNA ports (Write only) Saves SNP file corresponding to the entire active fixture. |
|
Parameters |
|
<snpName> |
String - snp file name. Appends ".sNp" depending on num ports |
Example |
CALC:FSIM:SAVE "myFixure" calculate:fsimulator:save "D:\myfixture" |
Return Type |
Not applicable |
Not applicable |
CALCulate<cnum>:FSIMulator:SECTion:CIRCuit:ENABle?
Applicable Models: E505xA with VNA ports (Read only) Returns all circuits ON/OFF for all ports. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
Examples |
CALC:FSIM:SECT:CIRC:ENAB? calculate2:fsimulator:section:circuit:enable? |
Query Syntax |
CALCulate<cnum>:FSIMulator:SECTionCIRCuit:ENABle? |
Return Type |
Boolean (1 = ON, 0 = OFF) |
ON |
CALCulate<cnum>:FSIMulator:SECTion:EXTension:ENABle?
Applicable Models: E505xA with VNA ports (Read only) Returns the port extensions ON/OFF for all ports. |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
Examples |
CALC:FSIM:SECT:EXT:ENAB? calculate2:fsimulator:section:extension:enable? |
Query Syntax |
CALCulate<cnum>:FSIMulator:SECTion:EXTension:ENABle? |
Return Type |
Boolean (1 = ON, 0 = OFF) |
OFF |
CALCulate<cnum>:FSIMulator:SECTion:ZCONversion:ENABle?
Applicable Models: E505xA with VNA ports (Read only) Returns all ZConversions ON/OFF (both SE and BAL). |
|
Parameters |
|
<cnum> |
Any existing channel number. If unspecified, value is set to 1 |
Examples |
CALC:FSIM:SECT:ZCON:ENAB? calculate2:fsimulator:section:zconversion:enable? |
Query Syntax |
CALCulate<cnum>:FSIMulator:SECTion:ZCONversion:ENABle? |
Return Type |
Boolean (1 = ON, 0 = OFF) |
ON |
Applicable Models: E505xA with VNA ports (Write only) Loads an active fixture from a file. |
|
Parameters |
|
<fileName> |
Fixture topology file name. |
Example |
CALC:FSIM:TOP:LOAD "example.topo" calculate:fsimulator:topology:load "example.topo" |
Return Type |
Not applicable |
Not applicable |
Applicable Models: E505xA with VNA ports (Write only) Saves the current active fixture topology to a file. |
|
Parameters |
Not applicable |
<fileName> |
Fixture topology file name. |
Example |
CALC:FSIM:TOP:SAVE "example.topo" calculate:fsimulator:topology:save "example.topo" |
Return Type |
Not applicable |
Not applicable |
CALCulate<cnum>:FSIMulator:ZCONversion:DIFFerential:STATe
Applicable Models: E505xA with VNA ports (Read Only) Returns the differential port impedance conversion function ON/OFF. |
|
Parameters |
|
<cnum> |
Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <cnum> is set to 1. |
Examples |
CALC:FSIM:ZCON:DIFF:STAT? |
Query Syntax |
CALCulate<cnum>:FSIMulator:ZCONversion:DIFFerential:STATe? |
Return Type |
Boolean 0 Differential port impedance conversion OFF 1 Differential port impedance conversion ON |
0 |
CALCulate<cnum>:FSIMulator:ZCONversion:DIFFerential:BPORt<pnum>:COMPLex
Applicable Models: E505xA with VNA ports (Read-Only) Returns the complex impedance value for the differential port impedance conversion function. |
|
Parameters |
|
<cnum> |
Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <cnum> is set to 1. |
<pnum> |
Balanced port number. Choose from 1 to 999. Note: See Balanced port versus Logical port. |
Examples |
CALC:FSIM:ZCON:DIFF:BPOR2:COMPL?
|
Query Syntax |
CALCulate<cnum>:FSIMulator:ZCONversion:DIFFerential:BPORt<pnum>:COMPlex? |
Return Type |
Numeric <real>,<img> Real part of the Impedance value in Units, Imaginary part of the Impedance value in Units. |
0,0 |
CALCulate<cnum>:FSIMulator:ZCONversion:DIFFerential:BPORt<pnum>:SCALar
Applicable Models: E505xA with VNA ports (Read Only) Returns the real part of impedance value for the differential port impedance conversion function. |
|
Parameters |
|
<cnum> |
Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <cnum> is set to 1. |
<pnum> |
Balanced port number. Choose from 1 to 999. Note: See Balanced port versus Logical port. |
Examples |
CALC:FSIM:ZCON:DIFF:BPOR2:SCAL
200,10 |
Query Syntax |
CALCulate<cnum>:FSIMulator:ZCONversion:DIFFerential:BPORt<pnum>:SCALer? |
Return Type |
Numeric Real part of the Impedance value in Units |
1 |
CALCulate<cnum>:FSIMulator:ZCONversion:DIFFerential:LPORt<pnum>:COMPLex
Applicable Models: E505xA with VNA ports (Read Only) Returns the complex impedance value for the common mode port impedance conversion function. |
|
Parameters |
|
<cnum> |
Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <cnum> is set to 1. |
<pnum> |
Logical port number. Choose from 1 to 999. Note: See Balanced port versus Logical port. |
Examples |
CALC:FSIM:ZCON:DiFF:LPOR2:COMPL? |
Query Syntax |
CALCulate<cnum>:FSIMulator:ZCONversion:DIFFerential:LPORt<pnum>:COMPlex? |
Return Type |
Numeric <real>,<img> Real part of the Impedance value in Units, Imaginary part of the Impedance value in Units |
0,0 |
CALCulate<cnum>:FSIMulator:ZCONversion:DIFFerential:LPORt<pnum>:SCALar
Applicable Models: E505xA with VNA ports (Read Write) Returns the real part of impedance value for the differential mode port impedance conversion function. See Critical Note |
|
Parameters |
|
<cnum> |
Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <cnum> is set to 1. |
<pnum> |
logical port number. Choose from 1 to 999. Note: See Balanced port versus Logical port. |
Examples |
CALC:FSIM:ZCON:DIFF:LPOR2:SCAL? |
Query Syntax |
CALCulate<cnum>:FSIMulator:ZCONversion:DIFFerential:LPORt<pnum>:SCALer? |
Return Type |
Numeric Real part of the Impedance value in Units |
1 |
CALCulate<cnum>:FSIMulator:ZCONversion:COMMonmode:STATe
Applicable Models: E505xA with VNA ports (Read Only) Returns the common mode port impedance conversion function ON/OFF. |
|
Parameters |
|
<cnum> |
Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <cnum> is set to 1. |
Examples |
CALC:FSIM:ZCON:COMM:STAT? |
Query Syntax |
CALCulate<cnum>:FSIMulator:COMMonmode:DIFFerential:STATe? |
Return Type |
Boolean 0 Common mode port impedance conversion OFF 1 Common mode port impedance conversion ON |
0 |
CALCulate<cnum>:FSIMulator:ZCONversion:COMMonmode:BPORt<pnum>:COMPLex
Applicable Models: E505xA with VNA ports (Read-Write) Returns the complex impedance value for the common mode port impedance conversion function. |
|
Parameters |
|
<cnum> |
Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <cnum> is set to 1. |
<pnum> |
Balanced port number. Choose from 1 to 999. Note: See Balanced port versus Logical port. |
Examples |
CALC:FSIM:ZCON:COMM:BPOR2:COMPL? |
Query Syntax |
CALCulate<cnum>:FSIMulator:ZCONversion:COMMonmode:BPORt<pnum>:COMPlex? |
Return Type |
Numeric <real>,<img> Real part of the Impedance value in Units, Imaginary part of the Impedance value in Units |
0,0 |
CALCulate<cnum>:FSIMulator:ZCONversion:COMMonmode:BPORt<pnum>:SCALar
Applicable Models: E505xA with VNA ports (Read Only) Returns the real part of impedance value for the common mode port impedance conversion function. |
|
Parameters |
|
<cnum> |
Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <cnum> is set to 1. |
<pnum> |
Balanced port number. Choose from 1 to 999. Note: See Balanced port versus Logical port. |
Examples |
CALC:FSIM:ZCON:COMM:BPOR2:SCAL? |
Query Syntax |
CALCulate<cnum>:FSIMulator:ZCONversion:COMMonmode:BPORt<pnum>:SCALer? |
Return Type |
Numeric Real part of the Impedance value in Units |
1 |
CALCulate<cnum>:FSIMulator:ZCONversion:COMMonmode:LPORt<pnum>:COMPLex
Applicable Models: E505xA with VNA ports (Read Only) Returns the complex impedance value for the common mode port impedance conversion function. |
|
Parameters |
|
<cnum> |
Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <cnum> is set to 1. |
<pnum> |
Logical port number. Choose from 1 to 999. Note: See Balanced port versus Logical port. |
Examples |
CALC:FSIM:ZCON:COMM:LPOR2:COMPL? |
Query Syntax |
CALCulate<cnum>:FSIMulator:ZCONversion:COMMonmode:LPORt<pnum>:COMPlex? |
Return Type |
Numeric <real>,<img> Real part of the Impedance value in Units, Imaginary part of the Impedance value in Units. |
0,0 |
CALCulate<cnum>:FSIMulator:ZCONversion:COMMonmode:LPORt<pnum>:SCALar
Applicable Models: E505xA with VNA ports (Read Only) Returns the real part of impedance value for the common mode port impedance conversion function. |
|
Parameters |
|
<cnum> |
Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <cnum> is set to 1. |
<pnum> |
logical port number. Choose from 1 to 999. Note: See Balanced port versus Logical port. |
Examples |
CALC:FSIM:ZCON:COMM:LPOR2:SCAL? |
Query Syntax |
CALCulate<cnum>:FSIMulator:ZCONversion:COMMonmode:LPORt<pnum>:SCALer? |
Return Type |
Numeric, Real part of the Impedance value in Units. |
1 |
CALCulate<cnum>:FSIMulator:ZCONversion:SENDed:PORt<pnum>:STATe
Applicable Models: E505xA with VNA ports (Read Only) Returns the single-ended impedance conversion function ON/OFF. Must also set the fixture simulator function to ON using CALC:FSIM:STAT. |
|
Parameters |
|
<cnum> |
Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <cnum> is set to 1. |
<pnum> |
Port number. Choose from 1 to 999. |
Examples |
CALC:FSIM:ZCON:SEND:PORT3:STAT? |
Query Syntax |
CALCulate<cnum>:FSIMulator:ZCONversion:SENDed:PORt:STATe? |
Return Type |
Boolean 0 single-ended port impedance conversion OFF 1 single-ended port impedance conversion ON |
1 |
CALCulate<cnum>:FSIMulator:ZCONversion:SENDed:PORt<pnum>:COMPLex
Applicable Models: E505xA with VNA ports (Read Only) Returns the complex impedance value for the single-ended port impedance conversion function. |
|
Parameters |
|
<cnum> |
Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <cnum> is set to 1. |
<pnum> |
Pport number. Choose from 1 to 999. |
Examples |
CALC:FSIM:ZCON:SEND:POR2:COMPL? |
Query Syntax |
CALCulate<cnum>:FSIMulator:ZCONversion:SENDed:PORt<pnum>:COMPlex? |
Return Type |
Numeric <real>, <img> Real part of the Impedance value in Units, Imaginary part of the Impedance value in Units. |
0,0 |
CALCulate<cnum>:FSIMulator:ZCONversion:SENDed:PORT<pnum>:SCALer
Applicable Models: E505xA with VNA ports (Read Only) Returns the real part of impedance value for the single-ended port impedance conversion function. See Critical Note |
|
Parameters |
|
<cnum> |
Channel number of the measurement. There must be a selected measurement on that channel. If unspecified, <cnum> is set to 1. |
<pnum> |
Pport number. Choose from 1 to 999. |
Examples |
CALC:FSIM:ZCON:SEND:PORT:SCAL? |
Query Syntax |
CALCulate<cnum>:FSIMulator:ZCONversion:SENDed:PORT<pnum>:SCALer? |
Return Type |
Numeric |
1 |