86113A Dual Optical Module
Starting with firmware release A.06.90, 86100D Compatibility mode is no longer being actively tested and supported.
The 86113A module incorporates two optical measurement channels. A single-mode optical source is used to perform the factory calibration. The module is a calibrated reference receiver that is measured to The module provides
- Two 2.85 GHz, integrated, calibrated optical channel with sensitivity to below –20 dBm (P avg). Wavelengths between 1000 nm and 1600 nm. A user calibration provides for consistent accuracy at any wavelength between 1000 nm and 1600 nm using an optical source and a precision optical power meter.
- Switchable reference filters for transceiver compliance testing.
- Compliance testing ITU-TS G.957 and TANWT-000253 frequency response requirements (Option 201) or to reference receiver specifications for Fibre Channel (FC) 1063 and Gigabit Ethernet for transmitter compliance testing (Option 202).
- Measurement capability for single mode or multimode optical signals
The minimum required 86100-series firmware version for this module is A.03.01.
See Also
- Available Accessories
- Optical Channel Calibration
- Modules
Supplied Accessories
Item | Description |
---|
Available Options
Item | Description |
---|---|
86113A-201 | OC-3/STM-1 (155 MBd), receiver filter OC-12/STM-4 (622 MBd), receiver filter |
86113A-202 | FC-1063 (1062.5 MBd), receiver filter Gigabit Ethernet (1250 MBd), receiver filter |
86113A-301 | OC-3/STM-1 (155 MBd), receiver filter OC-12/STM-4 (622 MBd), receiver filter OC-48/STM-16 (2.488 GBd), receiver filter |
86113A-UK6 | Commercial calibration certificate with test data |
81000AI | Diamond HMS-10 connector interface |
81000FI | FC/PC/SPC/APC connector interface |
81000KI | SC connector interface |
81000SI | DIN 47256/4108.6 connector interface |
81000VI | ST connector interface |