Last updated: January 15, 2009
This section is only applicable to the lab application.
This example uses an Keysight E4443A PSA Series Spectrum Analyzer.
Spectrum Analysis
.
Center Freq
to the appropriate "mid range" frequency for your UE's operating band, as specified by 3GPP TS 34.108 5.1.1 (for example, for operating band I,
1950.0 MHz
).
More
to move to the
2 of 3
menu. Set
REF Lvl OFFst
to
9 dB
to account for the loss due to the RF splitter (or to the appropriate value for your test setup).
More
to move to the
1 of 3
menu. Set
Ref Level
to
-30 dBm
(or to the appropriate level for your test setup to allow accurate measurement of the UE's off power).
Zero Span
.
Detector
to
Man
and choose
Average (Log/RMS/V)
(this also sets
BW/Avg
:
Avg/VBW Type
to
Pwr Avg (RMS)
).
Res BW
to
5 MHz
.
Ext Rear
or
Ext Front
.
RF IN/OUT Amplitude Offset
to
-9 dB
(or to the appropriate value for your test setup) to account for the loss due to the RF splitter (press the
SYSTEM CONFIG
key, then select
RF IN/OUT Amptd Offset
(
F5
)).
Cell Parameters.
Test Application: Press
Cell Parameters
(
F2
).
Cell Info
(
F2
), then
Cell Parameters
(
F2
).
Call Control 2 of 5
menu)
S200
.
15 s
.
S1
.
Generator Info
(
F3
). Select
Connected DL Channel Levels
(
F4
).
Measurement Generated
.
RB Test Mode
connection between the UE and test set (see
How Do I Set Up a Call and Make a Connection?
).
Call Parms 3 of 3
menu) to
28 dBm
.
UL CL Power Ctrl Parameters
(
F8
). Set
UL CL Power Control Mode
to
All Up bits
.
Call Parms 1 of 3
menu) to
-61 dBm
.
Generator Info
(
F3
on the
Call Control 2 of 5
menu) and set
AWGN Power
(
F5
) to
-60
dBm.
Out-of-Synch Handling of Output Power
.
Out-of-Sync Power Setup
and set the parameters as needed. The default values are those specified by 3GPP TS 34.121 5.4.4. Select
Close Menu
(
F6
).A typical result is shown below.
If desired, when you have completed the measurement, select
End Call
(
F3
) and return the following settings to their pre-measurement values:
Cell Parameters.
Test Application: Press
Cell Parameters
(
F2
).
Cell Info
(
F2
), then
Cell Parameters
(
F2
).
Call Control 2 of 5
menu)
Generator Info
(
F3
). Select
Connected DL Channel Levels
(
F4
).
Trig Output Setup
(
F6
) to
Sys Frame Clock
.
UL CL Power Ctrl Parameters
(
F8
). Set
UL CL Power Ctrl Mode
to
Active bits
.
UE Target Power
(
F7
on the
Call Parms 3 of 3
menu) as desired.
Cell Power
(
F7
on the
Call Parms 1 of 3
menu) as desired.
Generator Info
(
F3
on the
Call Control 2 of 5
menu) and set
AWGN Power
(
F5
) as desired.