Conditions to Monitor
Flex Apps:
FlexDCA
FlexRT
FlexPLL
This topic lists register bits that are associated with conditions that you want to monitor.
Condition | Status Name | Registers |
---|---|---|
Calibration completion | DONE | :STATus:CALibration:EVENt? (bit 0)
|
CAL | :STATus:OPERation:EVENt? (bit 14)
|
|
Clock recovery unlocked | UNLK | :STATus:CRECovery:EVENt? (bit 0))
|
CLCK | :STATus:OPERation:EVENt? (bit 7)
|
|
Clock recovery locked or trigger capture has occurred | LOCK | :STATus:CRECovery:EVENt? (bit 1)
|
CLCK | :STATus:OPERation:EVENt? (bit 7)
|
|
Clock recovery lock routine finished | FIN | :STATus:CRECovery:EVENt? (bit 6)
|
CLCK | :STATus:OPERation:EVENt? (bit 7)
|
|
FlexPLL transitioned from remote mode to local mode | LOCAL REG | :STATus:LOCal:EVENt? (bit 0)
|
LCL | :STATus:USER:EVENT? (bit 0)
|
|
LCL | *STB? (bit 1)
|
|
Message displayed and placed in message queue | MSG | :STATus:MESSage:EVENt? (bit 0)
|
MSG | *STB? (bit 2)
|
|
FlexPLL connected to N1000A | CON | :STATus:RDCA:EVENt? (bit 0)
|
RDCA | :STATus:OPERation:EVENt? (bit 6)
|
|
FlexPLL remote connection to N1000A timed out | TOUT | :STATus:RDCA:EVENt? (bit 1)
|
RDCA | :STATus:OPERation:EVENt? (bit 6)
|
|
FlexPLL not connected to remote N1000A | DCON | :STATus:RDCA:EVENt? (bit 2)
|
RDCA | :STATus:OPERation:EVENt? (bit 6)
|
|
Trigger has been received | TRG | :STATus:TRIGger:EVENt? (bit 0)
|
TRG | *STB? (bit 0)
|
|
An operation is complete | OPC | *ESR? (bit 0)
|
OPC | *STB? (bit 5)
|
|
Query error detected | QYE | *ESR? (bit 2)
|
QYE | *STB? (bit 5)
|
|
Device-dependent error detected | DDE | *ESR? (bit )
|
DDE | *STB? (bit 5)
|
|
Execution error detected | EXE | *ESR? (bit 4)
|
EXE | *STB? (bit 5) | |
Command error detected | CME | *ESR? (bit 5)
|
CME | *STB? (bit 5)
|
|
Off-to-on transition has occurred | PON | *ESR? (bit 7)
|
ESB | *STB? (bit 5)
|