Adjustments Overview

Below are the adjustments available in the N7870A application. Adjustments for the M9300A reference module are supported by the separate calibration application N7873A.

  • Running an adjustment will invalidate the calibration date of that module as well as the calibration information for all PXIe systems that the module belongs to.

  • The M9300A frequency reference module must be installed in slot 10 of the chassis.

  • In all test equipment configurations, cables and adapters must be properly torqued. Exceeding recommended torque values can cause damage to a cable or adapter, and can cause inaccurate test results.

 

Click on the links below for the adjustment list for the model being tested. Each topic page will show any applicable options that apply to the test.