Last updated: January 14, 2010
Initiates and fetches an audio frequency analyzer measurement as a sequential operation. Measurement results are returned in the following order: integrity, average audio level, average SINAD, average audio distortion. |
|
Query |
Integrity: see Integrity Indicator |
Initiates and fetches the fast device tune measurement as a sequential operation. Results are returned in the following order:
|
|
|
|
Test Application Revision: A.09 and above, with the required feature license |
|
OUTPUT 714;"READ:CFDTune:ALL?" |
Initiates and fetches a QPSK EVM measurement as a sequential operation. Measurement results are returned in the following order: integrity, EVM maximum, frequency error worst case, origin offset maximum, phase error maximum, magnitude error maximum, peak EVM maximum. If the multi-measurement state is on, each result represents the average of a number of QPSK EVM measurements. Multi-measurement state, by default, is off. To set up multi-measurements, see SETup:EVMQpsk:COUNt[:SNUMber] . |
|
Query |
Initiates and fetches an HSDPA block error ratio measurement as a sequential operation. Measurement results are returned in the following order: Integrity Indicator, Block Error Ratio in %, Information Bit Throughput in kbps, Positive Acknowledgments, Negative Acknowledgments, Statistical DTXs, Number of Blocks Tested, Median Channel Quality Indicator. |
|
Integrity Indicator - See Integrity Indicator . |
|
Test Application Revision: A.08 and above, with the required feature license |
Initiates and fetches a Reporting of Channel Quality Indicator measurement as a sequential operation. Measurement results are returned in the following order: Integrity Indicator, Reporting of Channel Quality Indicator Pass/Fail Result. |
|
Integrity Indicator - See Integrity Indicator . Reporting of Channel Quality Indicator
|
|
Initiates and fetches a swept audio measurement as a sequential operation. Measurement results are returned in the following order: integrity indicator, average swept audio level and average distortion for the first frequency point, average swept audio level and average distortion for the second frequency point, and so on. The number of frequency points is determined with the SETup:SAUDio:FREQuency:POINts command. For frequencies above 10 kHz, this measurement returns 9.91 E+37. |
|
Integrity indicator: see Integrity Indicator |
Queries (initiates and fetches) one spectrum monitor measurement as a sequential operation. Returns a comma separated list of 401 points representing the amplitude results of the trace data. |
|
Initiates and fetches a loopback block error ratio measurement as a sequential operation. Measurement results are returned in the following order: integrity, block error ratio, block error count, blocks tested, uplink missing blocks. |
|
Query |
|
OUTPUT 714;"READ:WBLerror?" |
Initiates and fetches a channel power measurement as a sequential operation. Measurement results are returned in the following order: integrity, average channel power. If the multi-measurement state is on, the result represents the average of a number of channel power measurements. Multi-measurement state, by default, is off. To set up multi-measurements, see SETup:WCPower:COUNt[:SNUMber] . |
|
Query | |
Initiates and fetches a change of TFC measurement as a sequential operation. Measurement results are returned in the following order: Integrity Indicator, Step Down Relative Power (average result if SETup:WCTFormat:COUNt[:SNUMber] is greater than 0), Step Down Error, Step Down Pass/Fail, Step Up Relative Power (average result if SETup:WCTFormat:COUNt[:SNUMber] is greater than 0), Step Up Error, Step Up Pass/Fail. |
|
Integrity Indicator - see Integrity Indicator .
|
|
OUTPUT 714;"READ:WCTFormat?" |
Initiates and fetches a dynamic power analysis measurement as a sequential operation. Measurement results are returned in the following order: integrity indicator, the number of steps measured, and a comma-separated list of average power values for each step measured (in dBm). |
|
Integrity Indicator - see Integrity Indicator . |
|
Test Application Revision: A.05 and above Lab Application Revision: B.01 and above Test Application Revision A.08 and Lab Application C.03, increased Number of Steps range to 0 to 200 (previously it was 0 to 100) |
|
OUTPUT 714;"READ:WDPAnalysis?" ! Returns integrity, number of steps measured, ! average power for each step |
Initiates and fetches an extended range dynamic power measurement as a sequential operation. Measurement results are returned in the following order: integrity indicator, the number of steps measured, and a comma-separated list of average power values for each step measured (in dBm). |
|
Integrity Indicator - See Integrity Indicator . |
|
OUTPUT 714;"READ:WEDPower?" ! Returns integrity, number of steps measured, ! average power for each step OUTPUT 714;"READ:WEDPower:STEPs?" ! Returns number of steps measured |
Initiates and fetches an occupied bandwidth measurement as a sequential operation. Measurement results are returned in the following order: integrity, average occupied bandwidth. If the multi-measurement state is on, the result represents the average of a number of channel power measurements. Multi-measurement state, by default, is off. To set up multi-measurements, see SETup:WOBWidth:COUNt[:SNUMber] . |
|
Query | |
Initiates and fetches a PRACH transmit on/off power measurement as a sequential operation. Measurement results are returned in the following order: Integrity, Pre Burst Off Power level, Pre Burst Off Power pass/fail, Burst On Power, Open Loop Power Error value, Open Loop Power Error pass/fail, Post Burst Off Power level, Post Burst Off Power pass/fail (see PRACH Transmit On/Off Power Measurement Results ). |
|
Query |
Open Loop Power Error pass/fail
Post Burst Off Power pass/fail
|
Initiates and fetches a phase discontinuity measurement as a sequential operation. Measurement results are returned as a comma separated list of values: Integrity Indicator, Number of Steps Measured, Worst Case Phase Discontinuity Step Number, Worst Case Phase Discontinuity in degrees, Worst Case EVM Step Number, Worst Case EVM in % rms. |
|
Integrity Indicator - see Integrity Indicator . Worst Case Phase Discontinuity Step Number |
|
OUTPUT 714;"READ:WPDiscon?" |
Initiates and fetches a perceptual evaluation of speech quality measurement as a sequential operation. Measurement results are returned in the following order: Integrity, Uplink PESQ Score, Downlink PESQ Score,PESQ Uplink Headroom, PESQ Downlink Headroom. |
|
Integrity Indicator - see Integrity Indicator . |
|
OUTPUT 714;"READ:WPESquality?" |
Initiates and fetches a PRACH preamble analysis measurement as a sequential operation. Measurement results are returned in the following order: Integrity Indicator, EVM (rms) in % rms, EVM (peak) in %, Frequency Error in Hz, Origin Offset in dB, Phase Error (rms) in degrees, Magnitude Error in %, Timing Error in chips, IQ Gain Imbalance in dB, IQ Phase Imbalance in degrees, Signature. |
|
Integrity Indicator - see Integrity Indicator . |
|
OUTPUT 714;"READ:WPPAnalysis?" ! Returns Integrity Indicator, EVM (rms) in % rms, EVM (peak) in %, ! Frequency Error in Hz, Origin Offset in dB, Phase Error (rms) in degrees, ! Magnitude Error in %, Timing Error in chips, IQ Gain Imbalance in dB, ! IQ Phase Imbalance in degrees, Signature |
Initiates and fetches a spectrum emissions mask measurement as a sequential operation. Measurement results are returned in the following order: integrity, overall pass/fail, in-channel power, and pass/fail results for each of the four frequency ranges. In-channel power is returned in dBm/3.84 MHz. The four frequency ranges for this measurement are as follows:
If the multi-measurement state is on, the result represents the average of a number of spectrum emissions measurements. Multi-measurement state, by default, is off. To set up multi-measurements, see SETup:WSEMask:COUNt[:SNUMber] . |
|
Query |
In-Channel Power (dBm/3.84 MHz)
|
Initiates and fetches a spectrum emissions mask measurement as a sequential operation. Measurement results are returned in the following order: integrity, overall pass/fail (0 = Pass, 1 = Fail). If the multi-measurement state is on, the result represents the average of a number of thermal power measurements. Multi-measurement state, by default, is off. To set up multi-measurements, see SETup:WSEMask:COUNt[:SNUMber] . |
|
Query |
|
Initiates and fetches a TX dynamic power measurement as a sequential operation. This query returns 101 comma-separated values, in the order of an integrity value, followed by 100 power results. This query always returns 101 power results, regardless of how many power measurements were performed by the test set when the measurement was executed. The number of power measurements performed by the test set is equal to the Number of Power Steps setting + 1, and can be queried using the FETCh:WTDPower:COUNt[:STEP]? command. If fewer than 100 power measurements were performed by the test set, the remaining power results are returned as 9.91 E+37. |
|
|
|
Initiates and fetches a thermal power measurement as a sequential operation. Measurement results are returned in the following order: integrity, thermal transmit power. If the multi-measurement state is on, the result represents the average of a number of thermal power measurements. Multi-measurement state, by default, is off. To set up multi-measurements, see SETup:WTPower:COUNt[:SNUMber] . |
|
Query | |
Initiates and fetches the waveform quality measurements as a sequential operation. Measurement results are returned in the following order: integrity, EVM maximum, frequency error worst case, origin offset maximum, phase error maximum, magnitude error maximum. If the multi-measurement state is on, each result represents the average of a number of waveform quality measurements. Multi-measurement state, by default, is off. To set up multi-measurements, see SETup:WWQuality:COUNt[:SNUMber] . |
|
Query | |
|
|