:LTESt Mask Limit Commands

Instrument:
N1010A
DCA-X
DCA-M
Meas. mode:
Scope
Eye
Jitter
TDR
Flex Apps:
FlexDCA
FlexRT

Creating a Eye Mask Limit Test

The following example shows how to create a limit test on eye mask number 3 and run the test until 100 failures have occurred. After the test completes, results are reported:

  • Waveform is saved to a file.
  • Summary file is created.
  • Monochrome screen image of the entire FlexDCA display is saved to a JPG file.

You can create a test for each active eye mask.

				:ACQuire:SINGle  //  Stop waveform acquisition.

				:LTESt:MTESt:MRESult3:FAILures 100  //  Create Test on Mask 3. Stop test after 100 failures.
				:LTESt:MTESt:SWAVeform:CHAN1A:STATe DISK  //  Save channel 1A data to file.
				:LTESt:MTESt:SWAVeform:CHAN1A:FNAMe "C:\Users\kentb\Documents\Keysight\FlexDCA\Waveforms\Mask Waveform.wfmx"  //  File Name.
				:LTESt:MTESt:SSUMmary:STATe ON  //  Save summary file.
				:LTESt:MTESt:SSUMmary:FNAMe "C:\Users\kentb\Documents\Keysight\FlexDCA\Limit Summaries\Mask Test 1.sum"  //  File Name.
				:LTESt:MTESt:SIMage:STATe ON  //  Save screen image.
				:LTESt:MTESt:SIMage:MONochrome ON  //  Monochrome image capture.
				:LTESt:MTESt:SIMage:WINDow DISPlay  //  Capture entire display.
				:LTESt:MTESt:SIMage:FNAMe "C:\Users\kentb\Documents\Keysight\FlexDCA\Screen Images\Mask.jpg"  //  File Name.
				:LTESt:MTESt:MRESult1:STATe ON  //  Turn limit test on.

				:ACQuire:RUN  //  Start waveform acquisition.
			*OPC?  Wait for test completion.

Test Setup

Report: Screen Image File

Report: Summary File

Report: Waveform File