:DISK Documentation Wizard Commands

Instrument:
DCA-X
DCA-M
Meas. mode:
Scope
Eye
Jitter
TDR
Flex Apps:
FlexDCA
FlexRT
FlexPLL

Configures and saves a Documentation Wizard file (.zip). Use the :DISK:DWIZard:FNAMe command to specify the name of saved file and the :DISK:DWIZard:SAVE command to actually save the file. To select the type of information that is saved in the zip file, use the commands shown in the following table.

Commands Used to Select Data to Save
Item Selected with Command Instrument Mode Saved in Zip File
Scope Eye TDR Jitter
:DWIZard:SSCReen command
(Screen Capture)
Screen capture of entire display (Entire UI Screen Capture.png)
:DWIZard:SSETup command
(Instrument Setup)
Instrument state file (InstrumentSetup.setx)
:DWIZard:SSINfo command
(System Information)
System information shown in the Help > About dialog (SystemInfo.txt)
Waveform Content Windows
:DWIZard:TIME:SDATa command
(Waveform Window Data)
  • Scope Mode: ASCII XY-value waveform files (example, Channel 1A (1A).txt)
  • Eye Mode: color-grade gray-scale database waveform files (example, Channel 1A (1A).cgsx)
  • Jitter Mode: color-grade gray-scale database waveform files (example, Channel 1A (1A).cgsx)
  • TDR Mode: ASCII XY-value waveform files for Ports (example, Port 1 (1A).txt)
  • All modes: screen capture of waveform content window (Waveform Window Screen Capture.png)
:DWIZard:T:OHMS:SDATa command
(Time-Ohms N Window Data)
  • T-parameter ASCII XY-value waveform file for each waveform displayed in a Time-Ohms waveform content window (example, T11 Ohms.txt)
  • Screen capture of waveform content window (Time-Ohms N Window Screen Capture.png)
:DWIZard:T:VOLtS:SDATa command
(Time-Volts N Window Data)
  • T-parameter ASCII XY-value waveform file for each waveform displayed in a Time-Volts waveform content window (example, T13 Volts.txt)
  • Screen capture of waveform content window (Time-Volts N Window Screen Capture.png)
:DWIZard:T:PERCent:SDATa command
(Time-% N Window Data)
  • T-parameter ASCII XY-value waveform file for each waveform displayed in a Time-% waveform content window (example, T21 Percent.txt)
  • Screen capture of waveform content window (Time-% N Window Screen Capture.png)
:DWIZard:MAGNitude:SDATa command
(Freq-Mag N Window Data)
    FFT function output
  • Screen capture of waveform content window (Freq-Mag N Window Screen Capture.png)
  • S-parameter ASCII XY-value waveform file for each waveform displayed in a Freq-Mag waveform content window (example, S21 Magnitude.txt)
  • Screen capture of waveform content window (Freq-Mag N Window Screen Capture.png)
:DWIZard:PHASe:SDATa command
(Phase N Window Data)
    FFT function output
  • Screen capture of waveform content window (Phase N Window Screen Capture.png)
  • S-parameter ASCII XY-value waveform file for each waveform displayed in a Phase waveform content window (example, S11 Phase.txt)
  • Screen capture of waveform content window (Phase N Window Screen Capture.png)
:DWIZard:GDELay:SDATa command
(GrpDelay N Window Data)
    FFT function output
  • Screen capture of waveform content window (GrpDelay N Window Screen Capture.png)
  • S-parameter ASCII XY-value waveform file for each waveform displayed in a GrpDelay waveform content window (example, S11 Delay.txt)
  • Screen capture of waveform content window (GrpDelay N Window Screen Capture.png)
:DWIZard:SSParam command
(S-Parameters)
Touchstone S-parameter file for a DUT setup (example, DUT N S-Parameters.s2p)
Jitter Mode
:DWIZard:JGRaphs:SDATa command
(Graphs Window Data)
  • Jitter database file (JitterDatabase.jdx). Can be loaded into jitter memory.
  • Screen capture of jitter graphs (Graphs Window Screen Capture.png)
:DWIZard:JITTer:SDATa command
(Jitter Window Data)
    Jitter results table
  • Screen capture of Jitter results table (Jitter Window Screen Capture.png)
:DWIZard:AMPLitude:SDATa command
(Amplitude Window Data)
  • Amplitude results table in a comma-delimited ASCII file (Amplitude Window Results.csv)
  • Screen capture of Amplitude results table (Amplitude Window Screen Capture.png)
Option JSA
:DWIZard:JSASpectrum:SDATa command
(JSA Spectrum Window Data)
    JSA memory
  • Screen capture of jitter spectrum window (JSA Spectrum Window Screen Capture.png)
:DWIZard:JSAResults:SDATa command
(JSA Results Window Data)
    JSA results table
  • Screen capture of JSA results table (JSA Results Window Screen Capture.png)
General Purpose
:DWIZard:MARKers:SDATa command
(Markers Window Data)
  • Marker results table as comma-delimited ASCII file (Markers Window Results.csv)
  • Screen capture of waveform with markers (Markers Window Screen Capture.png).
:DWIZard:MTESt:SDATa command
(Mask Test Window Data)
    Mask test results table
  • Screen capture of waveform with mask shown (Mask Test Screen Capture.png).
:DWIZard:HISTograms:SDATa command
(Histogram Window Data)
    Histogram results table
  • Screen capture of the waveform with histogram (Histograms Window Screen Capture.png)
:DWIZard:LLTResults:SDATa command
(Limit Line Window Data)
  • Limit Line results table in comma-delimited ASCII file (Limit Lines Window Results.csv)
  • Screen capture of waveform with limit lines shown (Limit Lines Screen Capture.png)
:DWIZard:RESults:SDATa command
(Results Window Data)
  • Measurement results table in comma-delimited ASCII file (Results Window Results.csv)
  • Screen capture of Results table (Results Window Screen Capture.png)

Example

:DISK:DWIZard:SSCReen ON
:DISK:DWIZard:SSETup ON
:DISK:DWIZard:SSINfo OFF
:DISK:DWIZard:RESults:SDATa ON
:DISK:DWIZard:FNAMe "%USER_DATA_DIR%\Documentation Results\DocumentationResults_PAM4.zip"
:DISK:DWIZard:SAVe
*OPC?